A free-standing
slimline console design providing an ISO-5/Class 100
atmosphere in an operator workstation environment.
Completely self-contained, it functions effectively in an
uncontrolled area. Individually tested and certified to meet
or exceed the specifications for Laminar Flow Clean-Air
Devices as per Institute of Environmental Science
IES-RP-CC-002-86.
Specifications
(see
details)
Filter - Three inch HEPA
(High Efficiency Particulate Air) type, aerosol challenged,
zero probed. Aluminum frame minipleat design, Minimum
efficiency of 99.99% at .3 microns.
Prefilter - 1" bonded
polyfiber media minimizes surface-loading. Quick-release
aluminum frame. Easily accessible from front of unit. 78 %
arrestance – 56-57 ASHRAE.
Airflow - Average measured
clean air velocity 90 fpm (+/-10 fmp). All measured values
within +/- 20% uniformity.
Construction - Woodboard
construction for minimum noise & vibration. White and blue
plastic laminate finish. Clear acrylic side panels for
visibility. Rear-loaded HEPA filter. Protective plastic egg
crate diffuser screen in front of HEPA filter keeps uniform
air velocity within work area.
Electrical & Power Requirements -
All electrical components are UL
Approved. Standard 115V, 15 Amp, 1 pH, 60 Hz. SJO power cord
w/ 3-prong ground plug.
NEC Electrical
Code
Motor/Blower Assembly -
High efficiency
motors to reduce electrical load. Motors are thermally
protected & self-lubricating. High capacity metal forward
curved blowers with direct drive motors. Fan statically
balanced to maintain constant airflow. Motors/Blowers are
selected to maintain constant required airflow rate up to 2x
the initial static pressure.
Lighting
- Recessed 2 pin
fluorescent lighting. 125 ft. Candle +/- 20 ft. Candles.
Plastic prismatic lens cover maximizes light diffusion.
Controls
- Solid state,
adjustable airflow control, 80-120 fpm, 15A max. load
capacity, Internally mounted to prohibit accidental change.
Illuminated switches on face panel for motor & lights. |